High flow humidification treatment instrument
The high-flow humidification treatment device is a treatment device that provides continuous positive pressure and heating and humidification. Air and oxygen are filtered by filter cotton and inhaled by the fan. After passing through the mixing chamber, the air-oxygen mixture is formed, and then output to the humidification. In the tank, the surface of the hot water is heated and humidified to reduce the stimulation of mechanical ventilation on the cardiopulmonary system, keep the alveoli moist, facilitate sputum suction and prevent airway obstruction, and control the heating of the heating plate by setting the target gas temperature value. This temperature has real-time monitor. After heating and humidification, the heating wire breathing circuit is used to output the patient through the nasal tube after the second temperature rise.
Applications
The high-flow respiratory humidification treatment instrument is used for patients with spontaneous breathing, and effective treatment is carried out by providing a certain flow, warming and humidifying breathing gas. These patients include patients undergoing humidification therapy, oxygen therapy, tracheal intubation, and tracheotomy. Cannot be used for life support equipment.
Features
● Three levels of temperature control, overheating (dry burning) protection, safe and effective.
● The measured values of flow, temperature, oxygen concentration and working status are displayed on the same screen at the same time.
● Flow rate automatic compensation function.
● The machine is equipped with high-density filter cotton to avoid cross-infection.
● The oxygen concentration can be adjusted in the whole range, no need to add an external oxygen flow meter.
● Integrated ultrasonic real-time oxygen concentration monitoring, oxygen-free battery consumption.
● Easy disinfection, only need to use common water-● soluble disinfectant to moisturize, soft cloth to clean the surface of the machine and the surface of the parts that may be contaminated.
Technical Parameter
Power Source:AC 220V / 110V
Supply Frequency:50Hz / 60Hz
Power Rate:220W ± 15%
Output Gas Temperature: Output temperature: 31℃ - 37℃ Tolerance : ± 2%
Level 1: 31℃ Level 2: 34 ℃ Level 3: 37 ℃
Heater Plate Over -temperature Cutout: 115 ℃ ± 5 ℃
Humidification Chamber: Auto-fill, maximum volume 240 ml
Oxygen Concentration Monitoring Range: 21% - 100%
Tolerance : ±5% ( ± 3% FS)
Flow range:2 - 80 L/min
Auditory alarm:≥55 dB (A)
Host
Host+circuit+Mobile Jacking
Accessories
adaptors
Heated chambers
Patient interface
> 3618med
> merchants recruiting net
>Made in China